An adjustable-rate user-printable rain gauge calibrator
Download
- Final revised paper (published on 16 Nov 2026)
- Preprint (discussion started on 09 Jan 2026)
Interactive discussion
Status: closed
Comment types: AC – author | RC – referee | CC – community | EC – editor | CEC – chief editor
| : Report abuse
-
RC1: 'Comment on egusphere-2025-5744', Anonymous Referee #1, 10 Apr 2026
- AC1: 'Reply on RC1', Matthew Tippett-Vannini, 12 May 2026
-
RC2: 'Comment on egusphere-2025-5744', Anonymous Referee #2, 20 Jul 2026
- AC2: 'Reply on RC2', Matthew Tippett-Vannini, 10 Aug 2026
Peer review completion
AR – Author's response | RR – Referee report | ED – Editor decision | EF – Editorial file upload
AR by Matthew Tippett-Vannini on behalf of the Authors (03 Aug 2026)
Author's response
Author's tracked changes
Manuscript
ED: Publish subject to minor revisions (review by editor) (07 Aug 2026) by Anette Eltner
AR by Matthew Tippett-Vannini on behalf of the Authors (10 Aug 2026)
Author's response
Author's tracked changes
Manuscript
ED: Publish as is (17 Aug 2026) by Anette Eltner
AR by Matthew Tippett-Vannini on behalf of the Authors (24 Aug 2026)
Author's response
Manuscript
In this manuscript, the authors present an adjustable‑rate, printable calibrator for field calibration of low‑cost rain gauges - a concept I find both elegant and meaningful. They describe modifications to a previous calibration device and report various test results for the modified calibrator with adjustable flow rates. Overall, the manuscript reads more like a technical note to me than a full research article; whether this fits the journal’s scope is, in my view, a decision for the editor(s). With this being said, I find the methods and results interesting and the manuscript in general suitable for publication. However, I have a few points for (minor to major) revision that should be addressed before final publication.
My main concern relates to the field evaluation in Section 4.7. Fist of all, it would be informative to specify the temporal resolution used for field calibration and whether the ATMOS 41 provides bucket‑tip counts per time interval or exact tip timestamps: Details on how this gauge is calibrated (presumably via a screw that adjusts the tipping point) could also be provided. Furthermore, demonstrating that the gauges in Kenya performed well while the one in the USA did not mainly shows that the latter was poorly maintained or malfunctioning, rather than providing a robust field validation of the AR3D. A more informative approach would be to test a set of well‑maintained, calibrated gauges under multiple predefined flow rates, thereby establishing the performance range under ideal conditions. This would be useful for identifying gauges in the field that require maintenance and/or recalibration. Extending such tests to other types of low‑cost gauges could also yield useful insights into the applicability and robustness of the AR3D.
Minor comments
- Sections 2 and 3 contain very short subsections and/or paragraphs, often limited to a single sentence or a table; these should be merged into longer, cohesive sections. Tables currently placed in Sections 2.3 and 2.4 could be moved in the Appendix.
- The “pyranometer cover” findings (4.6) do not require a separate subsection, this can be integrated into the results.
- Please check bibliographic entries for complete details and consistent citation formatting (including years).
Specific edits
- p. 1, line 13 - Convert the flow rates (ml min⁻¹) to equivalent rainfall rates.
- p. 3, line 1 - Replace “document” with “article” or “manuscript”.
- p. 6, line 141 - Use SI units consistently.